WGR Readings: 119 Items 
Timestamp
NodeID
Description
Reading
Units
LCL
UCL
Status
08/27/2026 02:56:393/1/4030/0/0/0_1st Flr: City Water Pressure68.9PSI4090OK
08/27/2026 03:16:294/1/5037/0/0/0_2nd Flr: AH Scrubber Signet pH0.7 7.59Lower Ctl Limit
08/27/2026 03:16:074/1/5031/0/0/0_2nd Flr: AH Scrubber Sump Pump Liquid Level1.0On/Off02OK
08/26/2026 20:48:394/1/5038/0/0/0_2nd Flr: AH1 Magnehelic 35% First Filter DP0.530InH2O01.5Inactive
08/27/2026 03:00:514/1/5039/0/0/0_2nd Flr: AH1 Magnehelic 95% Filter DP0.111InH2O01OK
08/27/2026 03:15:334/1/5042/0/0/0_2nd Flr: AH3 Magnehelic 10.064InH2O02OK
08/27/2026 03:10:454/1/5043/0/0/0_2nd Flr: AH3 Magnehelic 20.315InH2O02OK
08/14/2026 09:43:294/1/5030/0/0/0_2nd Flr: Cool Fogger WDR0.1 0100Inactive
08/27/2026 03:05:524/1/5032/0/0/0_2nd Flr: Rm 237 Fluid Level 11.0On/Off02OK
08/03/2026 12:30:514/1/5033/0/0/0_2nd Flr: Rm 237 Fluid Level 21.0On/Off02Inactive
08/27/2026 03:06:394/1/5035/0/0/0_2nd Flr: Rm 237 Light Sensor 20.0On/Off02OK
08/27/2026 03:07:284/1/5001/0/0/0_2nd Flr: Wet Scrubber Magnehelic1.54PSI0.81.75OK
08/24/2026 09:23:574/1/5036/0/0/0_2ND_Flr: AH Scrubber Signet Flow361.3PSI275800Inactive
08/26/2026 14:31:064/1/5034/0/0/0_ASML_TEST: Proc Gas Fault Alarm4.2VOLTS45Inactive
08/26/2026 14:31:064/1/5034/0/0/1_ASML_TEST: Proc Gas Shut Down Alarm0.0VOLTS45Inactive
08/27/2026 03:11:413/1/4033/0/0/0_CDA Flow (CFM): Equipment Pad101.1 0800OK
08/27/2026 03:14:133/1/4101/0/0/0_CDA Pressure115.2PSI90140OK
08/27/2026 03:08:510/1/1201/0/0/0_Cleanroom: Equip RH42% / 70.3F38 / 6542 / 75OK
08/27/2026 03:08:480/1/1054/0/0/0_Cleanroom: Rm 101C Rasco AC208V ON Light Indicator1.0On/Off02OK
08/27/2026 03:08:480/1/1054/0/0/1_Cleanroom: Rm 101C Rasco START Light Indicator1.0On/Off02OK
08/27/2026 02:19:174/1/5041/0/0/0_Cleanroom: Rm 101C_Argon_1050112583PSI2003000OK
08/27/2026 02:48:384/1/5040/0/0/0_Cleanroom: Rm 101C_Oxygen_105010368PSI2003000OK
08/27/2026 03:05:520/1/1055/0/0/0_Cleanroom: Rm 101D Haskris Building Water Fault Indicator0.0On/Off02OK
08/27/2026 03:17:233/1/4031/0/0/0_DI-RO: Light Sensor0.0On/Off02OK
08/27/2026 02:59:173/1/4017/0/0/0_DI-RO: Resistivity Gauge #1814.6Ohm-cm14100OK
08/27/2026 03:09:143/1/4019/0/0/0_DI-RO:DI Water Tank Pressure (Tank 2)6.44PSI410OK
08/27/2026 03:09:163/1/4018/0/0/0_DI-RO:RO Water Tank Pressure (Tank 1)8.20PSI410OK
08/27/2026 03:13:010/1/1065/0/0/0_GPN2 Pressure Gauge After Filters124.3PSI0400OK
08/27/2026 03:14:003/1/4015/0/0/0_House Vacuum-25.93inHg-30-23OK
08/27/2026 03:07:303/1/4016/0/0/0_House Vacuum: Water Circ Pump Temp67.7DEG F085OK
08/27/2026 03:04:233/1/4002/0/0/0_LN2_Tanks_LI-1-1208.9InH2O0300OK
08/27/2026 03:07:083/1/4004/0/0/0_LN2_Tanks_LI-1-2229.1InH2O0300OK
08/27/2026 03:03:413/1/4006/0/0/0_LN2_Tanks_LI-1-3117.0InH2O0300OK
08/27/2026 03:14:543/1/4001/0/0/0_LN2_Tanks_PI-1-1130.8PSI0300OK
08/27/2026 03:13:443/1/4003/0/0/0_LN2_Tanks_PI-1-258.1PSI0300OK
08/27/2026 03:03:173/1/4005/0/0/0_LN2_Tanks_PI-1-3150.7PSI0300OK
08/27/2026 03:13:490/1/1060/0/0/0_N2 Pressure Gauge Canon EX3 Stepper_Chase 121100.0PSI0110OK
08/27/2026 03:11:223/1/4020/0/0/0_PCW: Lower Temp Gauge44.3DEG F052OK
08/27/2026 03:11:493/1/4009/0/0/0_PCW: Middle Pressure Gauge82.4PSI6794OK
08/27/2026 03:12:163/1/4008/0/0/0_PCW: Upper Temp Gauge55.6F0160OK
08/27/2026 03:07:150/1/1061/0/0/0_Pressure Gauge At EMN2 Shut Off_East Wall115.6PSI0135OK
08/27/2026 03:10:353/1/4032/0/0/0_RO: Flow to 2361.95GPM0100OK
08/27/2026 03:08:390/1/1062/0/0/0_Tank 47 GPN2112.27PSI90400Error-28
08/27/2026 03:15:150/1/1064/0/0/0_Tank 47 UHPN2120.3PSI90400OK
08/27/2026 03:16:310/1/1063/0/0/0_UHPN2 Pressure Gauge After Filters114.656PSI0125OK
08/27/2026 03:10:130/1/1006/0/0/0C_AJA UHV Ebeam_122_Ar_1052882683PSI2002800OK
08/27/2026 03:10:070/1/1012/0/0/0C_AJACovaporator_129_Ar_1052491013PSI2002800OK
08/27/2026 03:08:180/1/1016/0/0/0C_Beneq ALD Deposition_132_Ar_1052502004PSI3002800OK
08/27/2026 03:14:500/1/1101/0/0/0C_Cabinet2_Gas Bunker_100% Silane_102089722PSI1501000OK
08/27/2026 03:08:050/1/1102/0/0/0C_Cabinet2_Gas Bunker_2% Silane_102089549PSI2003000OK
08/27/2026 03:03:320/1/1103/0/0/0C_Cabinet3_Gas Bunker_Ethylene_102090638PSI4003000OK
08/27/2026 03:08:320/1/1021/0/0/0C_DDC_133_CO2_105376820PSI3502800OK
08/27/2026 03:07:210/1/1022/0/0/0C_DDC_133_O2_1053752004PSI3002800OK
08/27/2026 03:15:540/1/1004/0/0/0C_EX3_121_F 0.95%/Kr 1.25%/Ne Balance_2023PSI3002800OK
08/27/2026 03:03:270/1/1003/0/0/0C_EX3_121_He_451PSI2002800OK
08/27/2026 03:17:310/1/1002/0/0/0C_EX6_102_He_NoPVS2085PSI2002800OK
08/27/2026 03:10:450/1/1001/0/0/0C_EX6_102_Kr 1.25%/Ne Balance_NoPVS253PSI3002600Lower Ctl Limit
08/27/2026 03:12:590/1/1014/0/0/0C_FC300 Flip Chip Bonder_132_4% H2/He_1052541564PSI3003000OK
08/27/2026 03:15:040/1/1009/0/0/0C_General Use ICP Etchers_126_Ar_1050081899PSI2002800Error-27
08/27/2026 03:05:590/1/1008/0/0/0C_General Use ICP Etchers_126_He_1050091840PSI2002800OK
08/27/2026 03:03:560/1/1023/0/0/0C_HummerSputteringSystem_137_Ar_105255991PSI2002800OK
08/27/2026 03:09:090/1/1011/0/0/0C_IR Group ICP Etcher_126_Ar_1052822084PSI2002800OK
08/27/2026 03:09:550/1/1010/0/0/0C_IR Group ICP Etcher_126_He_105284571PSI2002800OK
08/27/2026 03:13:060/1/1015/0/0/0C_Ontos 7 Oxide Removal_132_He_1052532137PSI2002800OK
08/27/2026 03:12:290/1/1018/0/0/0C_Oxford HD PECVD_133_Ar_1052691664PSI2002800OK
08/27/2026 03:05:400/1/1019/0/0/0C_Oxford HD PECVD_133_He_105270711PSI2002800OK
08/27/2026 03:11:550/1/1020/0/0/0C_Oxford HD PECVD_133_SF6_105271288PSI140350OK
08/27/2026 03:07:300/1/1013/0/0/0C_PlasmaTech Orange RIE_130_Ar_1052721647PSI2002800OK
08/27/2026 03:15:060/1/1106/0/0/0C_PlasmaTherm PECVD & High Density ICPCVD_138_Cylinder Pressure_N2O785PSI2003000OK
08/27/2026 03:10:000/1/1026/0/0/0C_PlasmaTherm PECVD & High Density ICPCVD_138_Line Pressure_N2O18PSI525OK
08/27/2026 03:12:290/1/1027/0/0/0C_PlasmaTherm PECVD_138_CF4/Oxygen 8% NOS_1052601831PSI3504000OK
08/27/2026 03:11:500/1/1028/0/0/0C_PlasmaTherm PECVD_138_He_1052591600PSI2004000OK
08/27/2026 03:08:570/1/1017/0/0/0C_RTP_121_N2/5% H2_NoPVS1717PSI5003000OK
08/27/2026 03:08:260/1/1007/0/0/0C_STJ Fluorine RIE & Apex ICP_125_Ar_105243877PSI2002800OK
08/27/2026 03:07:480/1/1029/0/0/0C_STS Deep Trench DRIE_138_Ar_1052661290PSI2004000OK
08/27/2026 03:11:320/1/1024/0/0/0C_STS Deep Trench DRIE_138_C4F8_10261320PSI5300OK
08/27/2026 03:08:140/1/1030/0/0/0C_STS Deep Trench DRIE_138_He_105265479PSI2004000OK
08/21/2026 05:29:420/1/1025/0/0/0C_STS Deep Trench DRIE_138_SF6_NoPVS335PSI2003000Inactive
08/27/2026 03:07:290/1/1035/0/0/0C_Supply_Pressure_152_Near_CHF3638PSI1501500OK
08/27/2026 03:00:450/1/1104/0/0/0C_SystemA_Gas Bunker_H2_102092974PSI5003000OK
08/27/2026 03:15:400/1/1105/0/0/0C_SystemB_Gas Bunker_H2_1020932099PSI5003000OK
08/27/2026 03:12:260/1/1056/0/0/1C_Weigh_Scale_122_Not_Connected_CH20.2LBS0300OK
08/27/2026 03:14:340/1/1041/0/0/1C_Weigh_Scale_125_CCL2F217.8LBS5300OK
08/27/2026 03:14:340/1/1041/0/0/0C_Weigh_Scale_125_CF411.2LBS3300OK
08/27/2026 03:06:200/1/1040/0/0/0C_Weigh_Scale_125_CHF330.2LBS3300OK
08/27/2026 03:06:200/1/1040/0/0/1C_Weigh_Scale_125_SF640.4LBS3300OK
08/27/2026 03:15:010/1/1051/0/0/1C_Weigh_Scale_126_Not_Connected0.0LBS0300OK
08/27/2026 03:15:010/1/1051/0/0/0C_Weigh_Scale_126_SF6112.1LBS3300OK
08/27/2026 03:15:590/1/1049/0/0/1C_Weigh_Scale_130_2nd CHANNEL NOT IN USE0.0LBS0300OK
08/27/2026 03:03:390/1/1050/0/0/0C_Weigh_Scale_130_CF420.8LBS3300OK
08/27/2026 03:03:390/1/1050/0/0/1C_Weigh_Scale_130_Not_Connected0.1LBS0300OK
08/27/2026 03:15:590/1/1049/0/0/0C_Weigh_Scale_130_SF6_Cobra Oxford ICP Etcher17.7LBS3300OK
08/27/2026 03:09:230/1/1046/0/0/0C_Weigh_Scale_138_C4F827.3LBS3300OK
08/27/2026 03:14:500/1/1042/0/0/0C_Weigh_Scale_138_Dichlorosilane12.0LBS5300OK
08/27/2026 03:04:200/1/1045/0/0/0C_Weigh_Scale_138_HBr35.8LBS220Upper Ctl Limit
08/27/2026 03:14:500/1/1042/0/0/1C_Weigh_Scale_138_Not_Connected0.3LBS0300OK
08/27/2026 03:09:230/1/1046/0/0/1C_Weigh_Scale_138_SF639.0LBS3300OK
08/27/2026 03:04:200/1/1045/0/0/1C_Weigh_Scale_138_T30_Not_Connected0.0LBS0300OK
08/27/2026 03:05:540/1/1053/0/0/1C_Weigh_Scale_143A_C4F89.7LBS3300OK
08/27/2026 03:05:540/1/1053/0/0/0C_Weigh_Scale_143A_SF631.5LBS3300OK
08/27/2026 03:17:370/1/1052/0/0/0C_Weigh_Scale_152_Far C4F837.7LBS3300OK
08/27/2026 03:16:350/1/1047/0/0/1C_Weigh_Scale_152_Far_CF423.7LBS3300OK
08/27/2026 03:16:350/1/1047/0/0/0C_Weigh_Scale_152_Far_CHF336.6LBS3300OK
08/27/2026 02:51:480/1/1044/0/0/1C_Weigh_Scale_152_Near_C4F828.9LBS3300OK
08/27/2026 03:16:370/1/1043/0/0/1C_Weigh_Scale_152_Near_CF445.3LBS3300OK
08/27/2026 03:16:370/1/1043/0/0/0C_Weigh_Scale_152_Near_CHF348.0LBS3300OK
08/27/2026 03:17:370/1/1052/0/0/1C_Weigh_Scale_152_Not_Connected0.2LBS0300OK
08/27/2026 03:08:030/1/1048/0/0/0C_Weigh_Scale_CH 130_BCL314.3LBS3300OK
08/27/2026 03:08:180/1/1039/0/0/0C_Weigh_Scale_CH 130_Chlorine68.4LBS5300OK
08/27/2026 03:12:260/1/1056/0/0/0C_Weigh_Scale_Chase 122_SAMCO RIE_CF4117.1LBS3300OK
08/27/2026 02:51:480/1/1044/0/0/0C_Weigh_Scale_Versaline DRIE_152_Near_SF652.4LBS3300OK
08/27/2026 03:08:030/1/1048/0/0/1CHANNEL NOT IN USE0.0LBS00OK
08/27/2026 03:08:180/1/1039/0/0/1CHANNEL NOT IN USE0.0LBS00OK
 3/1/4304/0/0/0Flow Meter_Not Installed Ch 1 GPM  No Data
 3/1/4304/0/0/1Flow Meter_Not Installed Ch 2 GPM  No Data
 0/1/1202/0/0/0High Precision Humidity Sensor_Not Installed mAMPS  No Data
 0/1/1202/0/0/1High Precision Temperature Sensor_Not Installed mAMPS  No Data
 0/1/1005/0/0/0Z_(not installed - interference) mAMPS  No Data
 0/1/1031/0/0/0Z_STS Deep Trench DRIE_138_4% NOS_Not_Installed    No Data